Data<voltage>:
Data format: integer
Data range: 50~5000
Data accuracy: 1
Data unit: V
--
Example:
Set the voltage for ACW test in STEP1 as 1000V.
Command message: FUNC:SOUR:STEP 1:AC:VOLT 1000
--Return Message:
Query message: FUNC:SOUR:STEP 1:AC:VOLT?, returns to the current ACW
voltage in STEP 1, such as 1000.
FUNC: SOURce:STEP:AC:UPPC
To set/inquiry about the UPPER current for ACW test.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:AC:UPPC<current>
Query message:
FUNC:SOUR:STEP <sn>:AC:UPPC?
--
Data<current>:
Data format: float
Data range: 0.001 ~20.000 mA
Data accuracy: 0.001mA
Data unit: mA
--
Example:
Set the upper current for ACW test in STEP1 as 1mA.
Command message: FUNC:SOUR:STEP 1:AC:UPPC 1
--Return Value:
Query message: FUNC:SOUR:STEP 1:AC:UPPC?
Return to the current limit of ACW in STEP 1, such as 1
FUNC
:
SOURce
:
STEP <sn>
:
AC
:
LOWC
To set /inquiry about the LOWER current for ACW test.
(
Syntax refer to
AC
:
UPPC)
FUNC
:
SOURce
:
STEP <sn>
:
AC
:
TTIM
To set /inquiry about the TEST time for ACW test.
--
Syntax:
Command message:FUNC
:
SOUR
:
STEP <sn>
:
AC
:
TTIM <time>
Query message: FUNC
:
SOUR
:
STEP <sn>
:
AC
:
TTIM?
--
Data<time>:
Data format: float
Data range: 0
~
999.9 (0 is OFF)
Data accuracy: 0.1s
Data unit: s
--
Example:
Set the TEST time for ACW test in STEP1 as 1s.
Command message: FUNC:SOUR:STEP 1:AC:TTIM 1
--Return Value: