13
operation.
3.1.7
File Function Key
It is used to save existing settings to the internal storage area of the instrument.
3.1.8
Numeric keypad
It is used to enter data for test parameters and modify the current set parameters.
3.1.9
Indicator Area
●
FAIL
In the test, when a test data exceeds the setting, the instrument judges that the test failed, FAIL
indicator is on.
The upper indicator is the overall "FAIL" indicator and the lower indicators are for the test units.
When any test unit indicates "FAIL", the overall test result is "FAIL".
●
PASS
After the test, no test data exceeding the initial setting was found, and the instrument judges that the
test passed, PASS indicator is on.
In the case of the test timing function is off (TIME OFF), the test can only end with ‘STOP’, no
PASS judgment.
The upper headlight is the overall "PASS" indicator and the lower indicators are for the test units.
When any test unit indicates "PASS", the overall test result is "PASS".
3.1.10
High voltage indication and Work instruction
This light is on when the instrument is ready for high voltage testing or when performing high
voltage testing.
Note: When the light is on, the operator should stay away from the high voltage tester, test
cable and the tested items to prevent electric shock.
3.1.11
Cursor control switch
The arrow keys are used to select the items to be modified, and the OK button is used to confirm
that the current project needs to be changed and the modification is completed.
3.1.12
Code switch and knob
Work together with cursor control switch to implement the parameters modification
3.1.13
Test function area (FUNCTION)
Select the working state of the instrument: including test status, test parameter setting status, and
system parameter setting status.
●
TEST
Press this button to light and the instrument enters the ready test state;
only in this state the
instrument is allowed to start the high voltage test.
●
SETUP
Press the button to light and the instrument enters the parameter setting interface;
only in this state
the instrument will modify the test parameters.