53
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:OS:STAND<capacitance >
Query message: FUNC:SOUR:STEP <sn>:OS: STAND?
--
Data< capacitance >:
Data format: float
Data range:
0.001 ~40.0nF
Data accuracy: 0.001nF
Data unit: nF
--
Example:
Set the standard capacitance for OS test in STEP1 to 1nF.
Command message: FUNC:SOUR:STEP 1:OS:STAND 1
--Return Message:
Query message: FUNC:SOUR:STEP 1:OS: STAND?
Return value: 1.000
FUNC
:
SOURce
:
STEP <sn>
:
OS
:
GET
Get current channel data as standard value.
Note: The instrument will initiate an OS test to get the current port value as a standard.
Please confirm that the port and test loop are configured correctly and pay attention to the
test safety.
Commands for CK Setup Function
FUNC
:
SOURce
:
STEP <sn>
:
CK
:
CH1
To set/ query CK scan box.
(There are only two states for CHx at CK.)
--
Syntax:
Command message: FUNC
:
SOUR
:
STEP <sn>
:
CK
:
CH1<scan type >
Query message: FUNC
:
SOUR
:
STEP <sn>
:
CK
:
CH1?
--
Data< scan type >:
Data format: character
Data range: ON/OFF(1/0)
--
Example:
Set CH1 of CK in STEP 1 to ON
Command message: FUNC
:
SOUR
:
STEP 1
:
CK
:
CH1 1
--Return Message:
Query message: FUNC
:
SOUR
:
STEP 1
:
CK
:
CH1?
Return value: ON
Note:
The CK function is used to check if the two leads are connected between the same test end.
The instrument automatically detects step by step when detecting multiple test ends.
Commands for DK Setup Function
FUNC
:
SOURce
:
STEP <sn>
:
DK
:
CH1
To set/ query DK scan box
--
Syntax:
Command message: FUNC
:
SOUR
:
STEP <sn>
:
DK
:
CH1<scan type >
Query message: FUNC
:
SOUR
:
STEP <sn>
:
DK
:
CH1?