36
4.3.6
D.CK Double-ended contact check parameter setting
It is valid to connect multi-channel scan box.
D.CK Function: Is the two test leads with two scan ports connected reliably? Generally connected to
the same pin of the DUT, it can only be used with the scan controller.
The setting interface is as follows:
Figure 4.3.6 D.CK setting interface
Contact Check (D.CK) the test parameters are as follows:
VOLT
:
0.50
~
0.500kV
High voltage test voltage value.
UPPER
:
0.1uA
~
5.00mA
Current upper limit.
SCAN
:
Red
This port is connected to high voltage for contact
inspection.
Green
This port is connected to the test end for contact
inspection.
Gray
This port is open.
UNIT
:
White
This unit participates in the test.
Gray
This unit does not participates in the test.
Description:
1.
The actual current is related to the internal resistance of the test loop. The default is the preferred
configuration
2.
The test status is DC power open loop mode, setting parameters and test values are for reference
only
3.
Can only achieve 1 pair of high and low end test at a time, after the test is over, the instrument
will automatically perform a rapid discharge of 0.2 seconds.
4.
Total Test Time=0.7S
4.4
Test function principle and instructions for use
This section takes the test process in order; introduce the principle and use of ground connection,
ground current detection, arc detection and other tests.