Procedure
1.
Select the “AUTO TEST”.
A window to confirm that the unit is rightly (ODS-D280U: horizontal/ODS-D280F: vertical) placed appears
waiting for entry of the [SET] button.
AUTO TEST
CHECK START?
PLACE DRIVE
HORIZONTALLY.
OK: SET KEY
EXIT: MENU KEY
AUTO TEST
CHECK START?
PLACE DRIVE
VERTICALLY.
OK: SET KEY
EXIT: MENU KEY
ODS-D280U
ODS-D280F
2.
Confirm that the unit is placed rightly and then click the [SET] button.
A confirmation window appears waiting for insertion of the working cartridge.
AUTO TEST
INSERT WORKING CTRG
EXIT: MENU KEY
OK: SET KEY
3.
Insert the working cartridge and the click the [SET] button.
The result is displayed.
AUTO TEST
CHECKING...OK
DISC 11/11
EXIT: MENU KEY
The following information is displayed when the check fails.
AUTO TEST
CHECKING...NG
EXIT: MENU KEY
DISC 09/11
6-4-5.
LOADER SWITCH
This menu is used to check the REC INH switch on the SM-48 board and door open/close sensor on the SM-49 board.
ODS-D280U/ODS-D280F
6-10