6-4-6.
SLIDER AUTO TEST
This menu is used to check operation of the optical block assembly and the limit sensor.
The following describes procedure for the optical block assembly OP-0. To check the optical block assembly OP-1,
select “OP-1” in step 2 below.
Note
When executing this menu, place the unit in the right position.
• ODS-D280U: horizontal
• ODS-D280F: vertical (right side surface below)
ODS-D280U: Horizontal
ODS-D280F: Vertical
If these sensors are checked with the unit placed improperly, the sensor check may fail.
Menu hierarchy
[CHECK]
[SLIDER]
[AUTO TEST]
Procedure
1.
Select the “AUTO TEST”.
A window to confirm that the unit is rightly (ODS-D280U: horizontal/ODS-D280F: vertical) placed appears
waiting for entry of the [SET] button.
SLIDER
CHECK START?
PLACE DRIVE
HORIZONTALLY.
OK: SET KEY
EXIT: MENU KEY
SLIDER
CHECK START?
PLACE DRIVE
VERTICALLY.
OK: SET KEY
EXIT: MENU KEY
ODS-D280U
ODS-D280F
2.
Confirm that the unit is placed rightly and then click the [SET] button.
A confirmation window appears waiting for insertion of the working cartridge.
SLIDER
INSERT WORKING CTRG
OK: SET KEY
EXIT: MENU KEY
ODS-D280U/ODS-D280F
6-13