54
C1.9 PLUS Ultrasound System Service Manual
12. Remove the four screws that secure the Trackball to the Control Panel
Subassembly using a #1 Phillips screwdriver (Figure 7.11).
Figure 7.11
Removing the Trackball
13. Holding the Trackball in place, carefully turn the Control Panel Subassembly
over and allow the Trackball to drop off of the Control Panel Subassembly, to
prevent losing the Trackball parts (Figure 7.12).
14. Remove the Trackball and note the location of its components.
Figure 7.12
Removing the Trackball
Screws (4)
Trackball
Trackball Wire
Harness
Summary of Contents for C1.9 PLUS
Page 1: ...C1 9 PLUS Ultrasound System Service Manual...
Page 4: ...iv C1 9 PLUS Ultrasound System Service Manual...
Page 12: ...4 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 18: ...10 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 38: ...30 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 50: ...42 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 72: ...64 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 88: ...80 C1 9 PLUS Ultrasound System Service Manual A 8 Accessory Cables 1 2 3 4 5 6 8 9 10 7 11...
Page 92: ...84 C1 9 PLUS Ultrasound System Service Manual...
Page 98: ...90 C1 9 PLUS Ultrasound System Service Manual...