48
C1.9 PLUS Ultrasound System Service Manual
7.2
Control Panel Subassembly Replacement
7.2.1
Required Parts
•
Service Assembly, Control Panel C1.9, English (P02082), or
•
Service Assembly, Control Panel C1.9, French (P02083), or
•
Service Assembly, Control Panel C1.9, German (P02084), or
•
Service Assembly, Control Panel C1.9, Italian (P02085), or
•
Service Assembly, Control Panel C1.9, Portuguese (P02086), or
•
Service Assembly, Control Panel C1.9, Spanish (P02087), or
•
Service Assembly, Control Panel C1.9, Japanese (P02544), or
•
Service Assembly, Trackball C1.9 (P02088)
Note:
The Service Assembly, Control Panel C1.9 does not include a trackball. To
install a trackball, see Section 7.3, Trackball, on page 52.
7.2.2
Required Tools
•
A #1 Phillips screwdriver, 7.0 in. (17.8 cm / 177.8 mm)
•
A torque driver, 2.0–10.0 in./lb (0.2–1.1 newton meter)
•
An anti-static mat
•
A wrist grounding strap
CAUTION:
`
Always use correct ESD procedures. ESD damage is cumulative and may not be
noticeable at first. Initial ESD symptoms may be slightly degraded performance
or image quality.
Summary of Contents for C1.9 PLUS
Page 1: ...C1 9 PLUS Ultrasound System Service Manual...
Page 4: ...iv C1 9 PLUS Ultrasound System Service Manual...
Page 12: ...4 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 18: ...10 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 38: ...30 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 50: ...42 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 72: ...64 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 88: ...80 C1 9 PLUS Ultrasound System Service Manual A 8 Accessory Cables 1 2 3 4 5 6 8 9 10 7 11...
Page 92: ...84 C1 9 PLUS Ultrasound System Service Manual...
Page 98: ...90 C1 9 PLUS Ultrasound System Service Manual...