Chapter 6: Troubleshooting 41
Figure 6.7
Battery Diagnosis
No Indication of
Power When You
Press the ON Button?
Did You Hear Two
Beeps When You
Pressed the Power
Button? Is There A
Video Display?
First Remove
Battery And
Then Connect
AC Power
Working?
Yes
No
Charge or
Replace
Battery
Yes
Yes
No
No Problem
Found
Insert
Battery and
Try Power
Again
No
No
Yes
No Problem
Found
Working?
Problem
Solved
Go To System
Diagram
Working?
Yes
No
Go To System
Diagram
Replace
Battery and
Try Again
Is Battery
Fully Inserted?