8. Remove the four screws that secure the Display Subassembly hinges to the top
housing with a #1 Phillips screwdriver (Figure 7.3).
9. Carefully guide the ends of the display wire harness and two 40-pin flex circuits
out through the slot in the top housing and remove the old Display Subassembly.
Figure 7.3
Removing the Display Subassembly
7.1.4
Replacing the Display Subassembly
1. Carefully guide the ends of the replacement display wire harness and two 40-pin
flex circuits through the slot in the top housing (Figure 7.4).
Figure 7.4
Replacing the Top Housing
Screws (4)
Top housing
Display hinges
Display Wire
Harness
40-pin flex
circuits (2)