Chapter 6: Troubleshooting 35
6.4.5
Clearing the Main PCBA Failure
After the assert code has been recorded, power down the system.
1.
Press the Power switch on the system until the power turns off (approximately
5–10 seconds).
2.
Turn the power back on to check if the fault cleared or if the condition remains.
If the condition cleared, you may use the system. If the condition remains,
corrective action must be taken before the system can be used.
6.4.6
Battery
If the system does not operate or does not run for the expected duration for a given
charge, battery failure is likely.