40
C1.9 PLUS Ultrasound System Service Manual
Figure 6.6
System Diagnosis
Battery
Charged?
Working?
Yes
Replace
Main PCB
Corrected?
Yes
No
No
No
Charge or
Replace
Battery
Try Battery
Power Only
Working?
Try AC
Power
Yes
No Problem
Found
Yes
Working?
No
Yes
Replace
AC
Adapter
No
Working?
System OK
Yes
Replace
Main PCB
No
Corrected?
Yes
Return
System To
SonoSite
No
No Indication Of
Power When You
Press The ON
Button?
Make Sure
Battery Is
Fully Inserted
Perform
Image Quality
Tests And
Verify AC
Power
Operation
Try AC
Power
If System Works On
Battery, Verify
Operation On AC
Power
First Remove
Battery And
Then Connect
AC Power
Perform
Image Quality
Tests And
Verify AC
Power
Operation
Listen For Two Beeps
And Look For A
Video Display.
Summary of Contents for C1.9 PLUS
Page 1: ...C1 9 PLUS Ultrasound System Service Manual...
Page 4: ...iv C1 9 PLUS Ultrasound System Service Manual...
Page 12: ...4 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 18: ...10 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 38: ...30 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 50: ...42 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 72: ...64 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 88: ...80 C1 9 PLUS Ultrasound System Service Manual A 8 Accessory Cables 1 2 3 4 5 6 8 9 10 7 11...
Page 92: ...84 C1 9 PLUS Ultrasound System Service Manual...
Page 98: ...90 C1 9 PLUS Ultrasound System Service Manual...