Chapter 7: Replacement Procedures 47
2. Secure the four screws on the replacement Display Subassembly hinges to the
top housing with a #1 Phillips screwdriver (Figure 7.3). Torque screws to
6.1 in./lb. (0.7 newton meter).
3. Place the Control Panel Subassembly onto the Display Subassembly.
4. Install the display wire harness and flex circuits.
Note: Check that the reference designators on the flex circuits match the reference
designators on the PCBA connectors.
5. Replace the 100-pin flex circuit on the bottom module (Figure 7.5).
Figure 7.5
Placing the Control Panel Subassembly onto the Display
6. Carefully replace the Control Panel Subassembly by inserting the tabs on the top
of the Control Panel Subassembly into the slots.
7. Set the Control Panel into place and replace the two screws inside the battery
compartment (Figure 7.6). Torque screws to 7.1 in./lb. (0.8 newton meter).
8. Place the battery in the battery compartment.
9. Turn on the system.
10. Perform the Display tests in
verify that the Display Subassembly is functioning properly.
Control Panel
Trackball
Display Wire
Harness
Anti-static mat
Display
100-pin flex
circuit
Summary of Contents for C1.9 PLUS
Page 1: ...C1 9 PLUS Ultrasound System Service Manual...
Page 4: ...iv C1 9 PLUS Ultrasound System Service Manual...
Page 12: ...4 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 18: ...10 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 38: ...30 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 50: ...42 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 72: ...64 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Page 88: ...80 C1 9 PLUS Ultrasound System Service Manual A 8 Accessory Cables 1 2 3 4 5 6 8 9 10 7 11...
Page 92: ...84 C1 9 PLUS Ultrasound System Service Manual...
Page 98: ...90 C1 9 PLUS Ultrasound System Service Manual...