PGA 3 Input of calibration values, Table 2 from 2
Step
Key
Function
8
Use the
numeric keypad
to enter a new calibration value (pulses x ml)
calibration value: SMD1A/1B:
1 Impuls x 2,6 ml SMD2:
1 Impuls x 9,3 ml
SMD3:
1 Impuls x 39 ml
SP/SMB 9/10: 1 Impuls x12 ml
SP/SMB13
1 Impuls x39 ml
SP/SMB14: 1 Impuls x 106 ml
SP/SMB15:
1 Impuls x 4,6 ml
Confirm input:
Press the
ENTER
key
Change another calibration-value measuring point:
Press
next step function key
Enter input as explained in previous description - step 8
Select data-save menu
Press
next step function key
until the data-save menu is jumped to
10
Store or reject calibration values
The data-save menu calls on the operator to send the new calibration values to the pulse meter for storage. When function key S1
is pressed the new calibration values are transferred to the IPM 12 pulse meter selected (in step 2) and stored.
Press
function key S1
(date save)
or reject new calibration values
Press
function key S2
The main menu is jumped back to without storage of the calibration data.
7. Operating
S1