3 Operation
Sirona Dental Systems GmbH
Analysis tools
Operator’s Manual SIDEXIS XG
Activating the measurement tool
➢ Click the button.
or
➢ In the menu bar
"[A]nalysis"
, select the
"Adapt [m]easurement"
function.
Adapting the length measurement
1. First measure a known reference length (e.g. a root canal file with defined
length) on the image.
2. Open the
"Adapt measurement"
dialog box.
ª
The previously measured value appears as
"Length"
in the dialog
box.
3. Enter the actual value as correction value in the
"Calculated length"
input
box. You can adjust this value in 0.1 mm steps using the two arrow keys.
ª
SIDEXIS XG displays the calculated correction factor.
4. Close the dialog box by clicking the
"OK"
button.
ª
For all subsequent length measurements in this image the calculated
correction factor is used.
ª
To make the user aware of the activated correction factor, all measured
values receive the label
"ref."
on the status bar and in the image.
Deactivating the correction
1. If further measurements are to be made without length correction, open
the
"Adapt measurement"
dialog box once again and activate the
"Don't
adapt"
option button.
2. Then close the dialog box by clicking the
"OK"
button.
Example
●
"Length"
= 11.26 mm
●
"Calculated length"
= 10,10 mm
●
"New factor"
= 0,897
All subsequent measurements in the same image are multiplied by 0.897.
Example screen
NOTICE
The correction values are sufficiently accurate only in the vicinity of the
reference value, because the angle of the sensor in relation to the tooth
may be different in other image areas.