10 Cephalometric exposures (CEPH)
Sirona Dental Systems GmbH
10.1 Preparing a cephalometric exposure (CEPH function)
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
59 87 594 D 3352
72
D 3352.201.01.18.02
10
Cephalometric exposures (CEPH)
10.1
Preparing a cephalometric exposure (CEPH function)
Select the CEPH function
by pressing
"CEPH"
in the program group selection.
The menu for program C3 lateral views (asymmetric A)
appears on the touchscreen.
Using the +/– arrow keys, you can select program C1
p.a. (symmetric S), C2 a.p. (symmetric S) or C4 for car-
pus views.
"R button, move into Ceph starting position"
appears in the comment line.
ATTENTION
Make sure that no patient is in the movement range of the
PAN rotating element or cephalometer.
Now press the
"R"
key on the Easypad.
The rotating element moves into the position for cepha-
lometric radiography. The secondary diaphragm and the
ceph sensor on the cephalometer move all the way to
the rear for patient positioning.
PAN
TS
9,1s
80kV
14mA
?
CEPH
C2 a.p.
C2 a.p.
1558
9,1s
Ready for exposure
PAN
9,1s
80kV
14mA
?
CEPH
C1 p.a.
C1 p.a.
1558
9,1s
TS
Ready for exposure
PAN
TS
9,1s
64kV
16mA
?
CEPH
C4
1260
9,1s
Ready for exposure
PAN
TS
73kV
15mA
?
CEPH
C3
1260
9,4s
9,4s
Ready for exposure