9 Operation
Sirona Dental Systems GmbH
9.1 Preparing the exposure
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
59 87 594 D 3352
52
D 3352.201.01.18.02
NOTE
i
After the
T
key is pressed, a
test cycle
of the rotating el-
ement can be triggered
without radiation
.
After the
T
key is pressed, the adjacent display appears
on the touchscreen without the kV/mA value, exposure
time and patient symbols. Two test cycle symbols appear
instead of the patient symbols.
To quit the test cycle mode, press the
T
key once again.
Switch SIDEXIS to ready for exposure state
•
Make the
SIDEXIS program on the PC
ready for
exposure
for PAN (XP)/CEPH (XC) or TS (XS) im-
ages. (see
SIDEXIS
, Operator's Manual).
As long as no connection with SIDEXIS is estab-
lished, the message
"Switch SIDEXIS to ready for
exposure state"
is displayed in the comment line of
the Easypad touchscreen.
Once SIDEXIS is ready for exposure, the welcome
screen with the selected patient data from
SIDEXIS appears on the Easypad touchscreen.
It shows the
first name, last name, date of birth and
card index number
of the patient currently registered in
SIDEXIS.
The right upper corner shows the radiation symbol, the
program number with the associated kV/mA values, and
a help message H.
When you touch the screen, the welcome screen disap-
pears and the selection screen reappears.
NOTE
i
If you wish to suppress this entire screen display or indi-
vidual pieces of information displayed there, your service
engineer can disable the corresponding data upon re-
quest.
Ready for exposure
P1
64kV
8mA
H301