1 Warning and safety information
Sirona Dental Systems GmbH
1.1 General safety information
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
59 87 594 D 3352
6
D 3352.201.01.18.02
Do not spray disinfectants or other similar products into
the ventilation slots.
X-raying of patients
X-rays of patients must be taken only when the system
works without errors.
The system may only be operated by skilled or prop-
erly trained personnel.
The movements of the unit must not be obstructed by
physical constitution nor clothing, dressings, wheel-
chairs or hospital beds!
Do not leave the patient unattended in the unit.
Electromagnetic compatibility (EMC)
Medical electrical equipment is subject to special pre-
cautionary measures regarding EMC. It must be
installed and operated as specified in the document
“Installation Requirements”.
Information on avoiding, detecting and eliminating unin-
tended electromagnetic effects: The
ORTHOPHOS
XG
Plus
DS/Ceph acquisition unit is Class
B
equipment
(classified according to CISPR 11, EN 60601-1-2: 2001
based on IEC 60601-1-2). This system may be operated
in a residential area.
Portable and mobile RF communications equipment
may interfere with medical electrical equipment. There-
fore, the use of mobile wireless phones in medical office
or hospital environments must be prohibited.
Please also observe the ESD protective measures in
Chapter 1.2.
Removing the sensor
To remove the sensor, hold it firmly, press the pushbutton
fully
in and hold it down. Remove the sensor from its
holder by pulling it downward.
DO NOT DROP THE SENSOR!
A shock sensor for detecting shocks or drops is built in.
When removing the sensor or handling a sensor that has
already been removed, make sure not to touch the sen-
sor plug on the unit end, especially not while touching
the patient at the same time.
Precautionary measures when switching on the
unit
Following extreme changes in temperature, condensa-
tion may occur; therefore, please do not switch the sys-
tem on until it has reached normal room temperature
(see chapter “Technical description”).
No patient may be positioned in the unit during
power-on.
In case of an error that requires switching the unit off and
back on again, the patient must be removed from the
unit, at the latest before switching the unit on again!