11 Transversal slices (TSA)
Sirona Dental Systems GmbH
11.6 TSA exposure with TSA slice orientation (alternative)
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
59 87 594 D 3352
102
D 3352.201.01.18.02
Startup settings menu
Level 4
You can access program level 3 by touching the disk icon
(21) in level 4.
In level 4 you can modify any factory-programmed
startup parameters.
They are then displayed after switching on the system
and for each new exposure.
Here you can change the patient icon preference (2nd
icon from the left, as per factory setting) and the slice
thickness preference (left icon for the thinnest slice, as
per factory setting).
Programming is done by touching the memory symbol
(1).
11.6
TSA exposure with TSA slice orientation (alternative)
During a TSA exposure, three slice views are taken.
Slices for anterior and canine teeth are 1 mm deep,
slices for the molar region are
approx. 2 mm deep.
You can preset the slices for the molar region to a slice
thickness of approx. 2 mm,
6 mm or 8 mm.
The slice gap is between 3 and 5 mm.
1.
Slice thickness (2, 6, 8 mm selectable)
2.
Slice center
3.
Slice gap
Angle preselection option
If the projection of the transversal slices is not orthogo-
nal to the path of the mandibular arch, the projection can
be adjusted with the angle preselection.
The TSA slices (3 x transversal) and (3 x trans 1
x lateral) allow for angle preselection (0°, 5°, 10°, 15°).
The angulation takes effect from the depth measure-
ment that applies to the molar range.
ATTENTION
The relationship of the angle to the slice orientation is
contra-rotatory
to that of the panoramic temporoman-
dibular joint programs TM1 and TM2.
TS
TS
Select Start Settings
1
0°
15°
1
.
3
.
2
.
Summary of Contents for Orthophos XG Plus DS/Ceph