13 List of messages
Sirona Dental Systems GmbH
13.2 Error message structure
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
59 87 594 D 3352
116
D 3352.201.01.18.02
13.2.3 yy
Digits (yy) define the location or logical function unit where the error has occurred.
The location may be a DX module number standing for an entire HW function unit or a logical SW function unit on the DX11
(central control).
13.2.4 zz
Digits (zz) show a consecutive number with the error ID.
5
Malfunction during the
exposure or during
exposure preparation.
Acknowledge the error message to
continue system operation. Repeat the last
procedure step or exposure. If the error
reoccurs -> contact your Customer
Service.
This error group includes all errors
resulting from a certain system action
triggered by the user which could not be
performed because a required (internal)
partial function (SW or HW) is not ready or
fails
6
Error during system
self-test.
Acknowledge the error message to con-
tinue system operation. If the error occurs
repeatedly, switch the system off and back
on; if the error reoccurs -> contact your
Customer Service.
This error group includes all errors which
may occur spontaneously and without any
related operator action. They may be
caused by system self-tests.
7
Unrecoverable sys-
tem error.
Switch off the system; immediately contact
your Customer Service.
This error group includes all errors which
may occur spontaneously and without any
related operator action. They may be
caused by system self-tests. In this case it
is absolutely certain that continued system
operation is not possible.
Ex
Description
Actions required
Error group
10
Central control DX 11; system hardware
11
Central control DX 11; system software
12
Central control DX 11; central CAN bus errors
13
Central control DX 11; DX11, DX1 periphery (motor system of stand, sensor system of stand)
14
Central control DX 11; digital extension (HSI, network...)
15
Central control DX 11; configuration (wrong software, wrong module constellation, etc.)
06
X-ray tube assembly
07
User interface (Comfort version/Easypad)
91
Ceph digital
81
Sensor
41
Media interface card
42
Remote
61
Diaphragm control
Summary of Contents for Orthophos XG Plus DS/Ceph
Page 1: ...b loqelmelp udm apL l f...