Sirona Dental Systems GmbH
9 Operation
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
9.4 Finishing the preparations (panoramic views)
59 87 594 D 3352
D 3352.201.01.18.02
65
Sinus views, programs S1 – S4 with
temporomandibular joint supports
•
For sinus views, you must fit the temporomandibular
joint supports (C)
"1"
on the right and (D)
"2"
on the
left side in place of the temple supports (7).
To do so, remove the two temple supports (7) after
pressing the corresponding locking button; instead,
push in the two temporomandibular joint supports
(C) and (D) until they engage.
Two sterile ear holders (E
1
) must be plugged into the
temporomandibular joint supports (C) and (D).
NOTE
i
For devices until October 2006:
Please observe the orientation (right/left) of the temporo-
mandibular joint supports specified on page 28.
9.4
Finishing the preparations (panoramic views)
•
If the light localizer is still on, switch it off with key
(28) on the Easypad. The light localizer symbol on
the touchscreen disappears.
•
The sensor (9) must be inserted up to the stop. This
is the case when the pushbutton (8) is flush with the
surface.
If the sensor (9) is not fully inserted, the help message
"Plug sensor into PAN slot"
appears in the comment
line.
If further help messages are displayed in the comment
line, they must be observed and processed one after the
other until the message "
Ready for exposure
" appears.
2
1
E
1
7
C
D
T
R
CEPH
PAN
TS
LS
Filmkassette einrasten
1260
64
62kV
8mA
?
P6.1
12,1s
CEPH
PAN
PAN
Aufnahmebereit
1260
64
64kV
8mA
?
P1
14,2 s
14,2 s
TS
8
9