6 Program group temporomandibular joint (TMJ) views
Sirona Dental Systems GmbH
6.2 TM2.1/TM2.2 Temporomandibular joints in posterior – anterior projection with closed and open mouth in one image
Operating
Instructions ORTHOPHOS XG
Plus
DS/Ceph
59 87 594 D 3352
40
D 3352.201.01.18.02
Program extensions:
TM2A.1 Temporomandibular in posterior - ante-
rior projection with closed mouth and TM2A.2
with open mouth, artifact-reduced
Touch the program display (6) to switch the exposure
type from TM2.1 or TM2.2 to TM2A.1 or TM2A.2 (arti-
fact-reduced).
•
Preparation as for TM2.1 or TM2.2.
•
Activate TM2A.1
.
On completion of
TM2A.1
the message “
Please
wait
" appears in the comment line and the unit auto-
matically returns to its starting position. The mes-
sage
"
Ready for exposure
” appears.
•
Ask the patient to open his/her mouth, and release
TM2A.2
.
•
Finally, the message “
R button, confirm exposure
data
” appears.
•
After you have pressed the R button again, the unit
returns to its starting position.
Angle preselection for temporomandibular
joint programs TM2.1/TM2.2 and TM2A.1/
TM2A.2
These programs allow for angle preselection
(0°, 5°, 10°, and 15°)
for the temporomandibular joint
area.
When you touch the angle symbol in the “submenu” col-
umn (8), a submenu line for angle preselection opens.
Select the desired angle; it will now be displayed in white
in column (8).
If detailed analyses of the temporomandibular joint are
required and the standard projections (0°) are not opti-
mal, the projections can be adjusted with the angle pre-
selection.
The figure shows the directions in which the slice orien-
tation is swiveled with angle preselection.
NOTE
i
When you confirm the exposure with the R button, the
angle setting that was changed in the submenu line will
automatically be reset to the default setting 0°.
0°
15°
AEC
CEPH
PAN
PAN
1260
10
71kV
8mA
?
TM2.1
18,7s
TS
10
12,8s
15°
1260
1260
15°
0°
5°
10°
Ready for exposure
Ready for exposure
18,7s
1
TS
0°
TM2A.2
Ready for exposure
AEC
CEPH
PAN
PAN
1260
10
71kV
8mA
?
TM2A.1
18,7s
TS
0°
Ready for exposure
6
8
TM2A.1
TM2A.2