MultiXplo / MultiTox
Page 40 / 60
NOSP15759-Rev 10
Calibration
Calibration must be made with the gas the detector is set to detect, with the SET menu (F2
key) of the wireless communication tool TLU600 and a calibration kit. The calibration gas
should be injected at a flow rate between 30 l/h and 60 l/h.
The detector is in maintenance inhibition mode for 10 minutes after it goes back to main
screen. Use the setting menu and validate the INH command for acknowledgement.
To end of communication between the TLU600 and the detector is done by pressing F4 key on
main menu.
Summary of Contents for DM-TT6-K
Page 9: ...MultiXplo MultiTox Page 9 60 NOSP15759 Rev 10 Figure 1 Layout Drawing Congestion see Figure 2...
Page 32: ...MultiXplo MultiTox Page 32 60 NOSP15759 Rev 10...
Page 48: ...MultiXplo MultiTox Page 48 60 NOSP15759 Rev 10...
Page 56: ...MultiXplo MultiTox Page 56 60 NOSP15759 Rev 10 10 CONFORMITY CERTIFICATE...
Page 57: ...MultiXplo MultiTox Page 57 60 NOSP15759 Rev 10...
Page 58: ...MultiXplo MultiTox Page 58 60 NOSP15759 Rev 10...
Page 59: ...MultiXplo MultiTox Page 59 60 NOSP15759 Rev 10...