3-30K
Malfunctions and Error Correction
62
Version 12/2008, Rev. 1.11 of 27/01/2012
Translation of the original operating manual
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a
centrifuge run, the centrifuge will be switched off immediately and
decelerate in an unbraked manner.
7.2 Error Codes
Error no.
Kind of error
Measure
Note
1-9
System error
allow to slow down
power off/on
All these errors stop the
centrifuge or cause it to
decelerate brakeless
10-19
Speedometer error
allow to slow down
power off/on
20-29
Motor error
power off/on
ensure ventilation
30-39
EEPROM error
allow to slow down
power off/on
With error 34, 35, 36, the
centrifuge will stop;
with error 37, 38 error
message only
40-45
Temperature error
(only for refrigerated
centrifuges)
allow to slow down
power off/on
allow to cool down
provide better ventilation
46-49
Imbalance error
allow to slow down
power off
eliminate the imbalance
50-59
Lid error
press lid key
close lid
remove foreign matter from the
opening of the lid lock device
With error 50 and 51, the
centrifuge will stop
60-69
Process error
allow to slow down
power off/on
With error 60 message
"power failure during run",
with error 61, message "stop
after power on"
70-79
Communication error
allow to slow down
power off/on
80-89
Parameter error
allow to slow down
power off/on
provide for better ventilation
With error 83, error
message only
90-99
Other errors
check connections
Fig. 7.4: Error codes
If it is impossible to eliminate the errors, contact the service (see 7.3)!