Spincontrol Comfort – Using the Centrifuge
Version 12/2008, Rev. 1.11 of 27/01/2012
39
Translation of the original operating manual
6.3.3.11 Modification of Parameter Values during the Centrifugation Run
(see 6.3.3.4 “Selection, Display, and Modification of Data"):
Modification possible Modification
impossible
(De-)Activation
possible
Speed
Rotor
Continuous run
RCF
Program
Standstill cooling
Runtime
Radius
Start delay
Temperature
Density
Automatic lid opening
function
Acceleration curve
Deceleration curve
6.3.4 Program Mode
A program includes all of the data that are required for a
centrifugation run.
Attention: Temperature profiles cannot be created!
With a program, certain sedimentation results can be repeated under
identical conditions.
Programs can be loaded, executed, modified, and deleted at any
time.
A maximum of 50 programs can be stored under the numbers 1 - 50.
The rapid cooling program "RAPID_TEMP" does not occupy any
storage location and cannot be delted. It is used to cool the centrifuge
without vessels.
"--" means that the values that are currently set are not a stored
program.
The programs can be protected against unauthorised use,
modification, or deletion with the aid of a code (see 6.3.3.7
"Configuration Menu" ).