![Sigma 3-30K Operating Manual Download Page 36](http://html1.mh-extra.com/html/sigma/3-30k/3-30k_operating-manual_1267902036.webp)
3-30K
Using the Centrifuge – Spincontrol Comfort
36
Version 12/2008, Rev. 1.11 of 27/01/2012
Translation of the original operating manual
Cycles
For each rotor/bucket combination, the number of cycles and runtime
are stored.
Buzzer
In this menu, a sound signal can be selected for
the end of a centrifugation run
an imbalance message
an error message.
The duration of the buzzer signal can be specified.
Sensor
The sensor overview shows various system data. In the event of a
malfunction, these data facilitate an immediate diagnosis.
Info
The Info overview shows information such as the type of centrifuge,
and EPROM version, number of cycles, the total runtime, as well as
software version and date.
In the menus "Cycles", “Info“, and “Sensor“, the values can neither be
entered nor changed.
Reset
This function resets the centrifuge and restores the factory settings.
All of the programs, parameters, curves, and configurations will be
deleted.
Curves
In this submenu, the user-defined acceleration and deceleration
curves can be created and edited (see 6.3.3.8).
6.3.3.8 Curves Menu
The Curves menu is a submenu of the configuration menu. The
curves function can be used to create user-specific acceleration and
deceleration curves. 10 storage locations (no. 20-29) are available for
this purpose.
For the interval no. 1, either a linear (LIN) or quadratic (QUAD)
acceleration can be selected. The other acceleration curves are
always linear.