![Sigma 3-30K Operating Manual Download Page 57](http://html1.mh-extra.com/html/sigma/3-30k/3-30k_operating-manual_1267902057.webp)
3-30K
Spincontrol Professional – Using the Centrifuge
Version 12/2008, Rev. 1.11 of 27/01/2012
57
Translation of the original operating manual
6.4.4 Program Mode
A program contains all data that are required for a centrifuge run.
Attention: Temperature profiles cannot be created!
With a program, certain sedimentation results can be repeated under
identical conditions.
Programs can be loaded, executed, modified, and deleted at any
time.
A maximum of 60 programs can be stored under the numbers 1 - 60.
The rapid cooling program "RAPID_TEMP" does not occupy any
storage location and cannot be delted. It is used to cool the centrifuge
without vessels.
"--" means that the values that are currently set are not a stored
program.
The programs can be protected against unauthorised use,
modification, or deletion with the aid of a code (see 6.4.3.7
"Parameter Menu").
6.4.4.1 Loading a Program
Select the option "Progr" from the “Standard” menu and confirm
the selection. The program selection list will be displayed.
Select the desired program from the list and confirm the selection.
The program is now loaded.
Fig. 6.39: Loading a program