SIGLENT Remote Manual
79 SDM3055/SDM3055A
2.9.27 [SENSe:]VOLTage:{AC|DC}:NULL:VALue:AUTO {ON|1|OFF|0}
[SENSe:]VOLTage:{AC|DC}:NULL:VALue:AUTO?
Enables or disables automatic null value selection for AC voltage or DC voltage and ratio
measurements.
Note:This parameter is not shared between AC and DC measurements. The parameters
are independent for AC and DC measurements.
Parameter
Typical Return
{ON|1|OFF|0}
,
default ON
0OFF)or1(ON)
Make a second set of measurements using automatic null value
selection:VOLT:AC:NULL:VAL:AUTO ON
READ?
Typical Response
:
+0.0000,+0.0100
When automatic reference selection is ON, the first measurement made is used as
the null value for all subsequent measurements.
[SENSe:]VOLTage:{AC|DC}:NULL:VALue will be set to this value. Automatic null
value selection will be disabled.
When automatic null value selection is disabled (OFF), the null value is specified by
this command:[SENSe:]VOLTage:{AC|DC}:NULL:VALue.
The instrument enables automatic null value selection when the null function is
enabled ([SENSe:]VOLTage: {AC|DC}:NULL:STATe ON).
This parameter is set to its default value after a Factory Reset (*RST), Instrument
Preset (SYSTem:PRESet), or CONFigure function.