Input terminal protection limitation
Protection limitation is defined for the input terminal
:
1. Main input
(
HI and LO
)
terminal
HI
and
LO
terminals are used for Voltage, Resistance,
Capacitance, Continuity, Frequency, Diode and temperture
measurement. Two protection limitations are defined:
HI-LO protection limitation:
1000VDC or 750AVC. It’s
the maximum measurable voltage. The limitation can be
expressed as 1000Vpk.
LO-ground protection limitation
:
LO
terminal can “float”
500Vpk relative to the ground safely. The maximum
protection limitation of
HI
terminal relative to the ground is
1000Vpk. Therefore, the sum of the “float” voltage and the
measured voltage can’t exceed 1000Vpk.
2. Sampling (HI
Sense
and LO
Sense
) terminal
HI
Sense and
LO
Sense are used for 4-wire Resistance
measurement. Two protection limitations are defined:
HI
Sense
-LO
Sense
protection limitation:
2000Vpk.
LO
Sense
-LO
Sense
protection limitation:
2Vpk.
3. Current input (I) terminal
I and LO terminals
are used for current measurement. The
maximum current which go through the
I
terminal
is limited to
10A by the fuse on the back panel.
NOTE:
Voltage on the current input terminal corresponds to voltage
on
LO
terminal. To keep good protection, only use the fuse of
specified type and level to replace this fuse.
Summary of Contents for SDM3055A
Page 1: ...User Manual SDM3055 Digital Multimeter UM06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Page 101: ...SIGLENT SDM3055 Digital Multimeter 91 Diagram 3 18 Result 2 ...
Page 112: ...Quick Strat SDM3055 Digital Multimeter QS06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Page 113: ......
Page 123: ...SIGLENT 2 SDM3055 Quick Start Appearance and Size Front View Side View 260 mm 105 mm 282 mm ...