2.9.5 [SENSe:]CURRent:{AC|DC}:RANGe:AUTO {OFF|ON|ONCE}
[SENSe:]CURRent:{AC|DC}:RANGe:AUTO?
Disables or enables autoranging for AC or DC current measurements.
Note:This parameter is not shared between AC and DC measurements. The parameters
are independent for AC and DC measurements.
Parameter
Typical Return
{ON|1|OFF|0}
,
default ON
0 (OFF) or 1 (ON)
Configure AC current measurements and perform an immediate autorange.
Make and read two measurements:
CONF:CURR:AC
CURR:AC:RANG:AUTO ONCE
SAMP:COUN 2
READ?
Typical Response
:
+5.79294185E-06,+5.79294185E-06
With autoranging enabled, the instrument selects the range based on the input signal.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument
Preset (SYSTem:PRESet).
2.9.6 [SENSe:]CURRent[:DC]:NPLC {<PLC>|MIN|MAX|DEF}
[SENSe:]CURRent[:DC]:NPLC? [{MIN|MAX|DEF}]
Sets the integration time in number of power line cycles (PLC) for DC current
measurements. Integration time is the period that the instrument's analog-to-digital (A/D)
converter samples the input signal for a measurement.
Parameter
Typical Return
{0.3|1|10}
,
default 1
On the front panel, 0.3|1|10 corresponds to the Speed
menu under fast|middle|slow
+1.0001
Configure DC current measurements using a 10 PLC integration time.:
CONF:CURR:DC
CURR:DC:NPLC 10
This parameter is set to its default value after a Factory Reset (*RST) or Instrument
Preset (SYSTem:PRESet).
Summary of Contents for SDM3055A
Page 1: ...User Manual SDM3055 Digital Multimeter UM06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Page 101: ...SIGLENT SDM3055 Digital Multimeter 91 Diagram 3 18 Result 2 ...
Page 112: ...Quick Strat SDM3055 Digital Multimeter QS06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Page 113: ......
Page 123: ...SIGLENT 2 SDM3055 Quick Start Appearance and Size Front View Side View 260 mm 105 mm 282 mm ...