2.9.30 [SENSe:]VOLTage[:DC]:NPLC {<PLC>|MIN|MAX|DEF}
[SENSe:]VOLTage[:DC]:NPLC? [{MIN|MAX|DEF}]
Sets the integration time in number of power line cycles (PLC) for DC voltage and ratio
measurements. Integration time is the period that the instrument's analog-to-digital (A/D)
converter samples the input signal for a measurement. A longer integration time gives
better measurement resolution but slower measurement speed.
Parameter
Typical Return
{0.3|1|10}
,
default 1
On the front panel, 0.3|1|10 corresponds to the
Speed menu under fast|middle|slow
+1.0001
Configure DC voltage measurements using a 10 PLC integration time.
CONF:VOLT:DC
VOLT:DC:NPLC 10
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).