2.10 SYSTem Subsystem
2.10.1 SYSTem:BEEPer:STATe {ON|1|OFF|0}
SYSTem:BEEPer:STATe?
Disables or enables the beep heard during continuity, diode, or Probe Hold measurements,
or when an error is generated from the front panel or remote interface.
Parameter
Typical Return
{ON|1|OFF|0}
,
default ON
0(OFF)or1(ON)
Disable keyboard click:
SYST:BEEP:STAT OFF
This command not affect the beeper heard associated with diode and continuity
tests, errors and so on.
This setting is non-volatile; it is not changed by power cycling, a Factory Reset
(*RST), or an Instrument Preset (SYSTem:PRESet).
2.10.2 SYSTem:COMMunicate:LAN:IPADdress "<address>"
SYSTem:COMMunicate:LAN:IPADdress? [{CURRent|STATic}]]
Assigns a static Internet Protocol (IP) address for the instrument.
Parameter
Typical Return
Command
:
"nnn.nnn.nnn.nnn"
Query
:
{CURRent|STATic}
,
default CURRent
Default
:
169.254.3.5
“192.168.1.101”
Set a static IP address:
SYST:COMM:LAN:IPAD "10.11.13.212"
This setting is non-volatile; it is not changed by power cycling, a Factory Reset
(*RST), or an Instrument Preset.(SYSTem:PRESet).
This parameter is set to its default value when the instrument is shipped from the
factory and after SYSTem:SECurity:IMMediate.
Summary of Contents for SDM3055A
Page 1: ...User Manual SDM3055 Digital Multimeter UM06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Page 101: ...SIGLENT SDM3055 Digital Multimeter 91 Diagram 3 18 Result 2 ...
Page 112: ...Quick Strat SDM3055 Digital Multimeter QS06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Page 113: ......
Page 123: ...SIGLENT 2 SDM3055 Quick Start Appearance and Size Front View Side View 260 mm 105 mm 282 mm ...