2.9.11 [SENSe:]{FREQuency|PERiod}:VOLTage:RANGe
{<range>|MIN|MAX|DEF}
[SENSe:]{FREQuency|PERiod}:VOLTage:RANGe? [{MIN|MAX|DEF}]
Manually choose frequency/cycle measurement mode voltage range:
Note
:
This parameter is shared between frequency and period measurements.
Parameter
Typical Return
<range>
:
{200 mV|2 V|20 V|200 V|750V}
,
default20V
+2.0001
Configures frequency measurements using the 10 VAC range. Make and read two
measurements:CONF:FREQ
FREQ:VOLT:RANG 20
SAMP:COUN 2
READ?
Selecting a fixed range ([SENSe:]<
function
>:RANGe) disables autoranging.
If the input voltage is too large for the selected voltage range (manual ranging), the
instrument displays the word Overload on the front panel and returns "9.9E37" from
the remote interface. Autoranging can be enabled for the input voltage.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument
Preset (SYSTem:PRESet).