53 SDM3055/SDM3055A
2.8.8 MEASure:CAPacitance [{<range>|AUTO|MIN|MAX|DEF}]
Sets all measurement parameters and trigger parameters to their default values for
capacitance measurement.Also specifies the stalls through the incoming parameters
.
Parameter
Typical Return
<range>
:
{2nF|20nF|200nF|2uF |20uF|
200uF|10000uF}
,
AUTO (default) or DEFault
(
none
)
Configure DC voltage ratio measurements using the 2uF resolution. Then make
and read one
measurement:
MEAS:CAP? 2uF
READ?
Typical Response
:
+7.26141264E-10
For the
Input
terminals, you can allow the instrument to select the measurement
range by autoranging or you can select a fixed range using
manual ranging
.
Autoranging decides which range to use for each measurement based on the input
signal. For faster measurements, use manual ranging on each measurement
(additional time is required for autoranging to select a range).
If the input signal is greater than can be measured on the specified manual range, the
instrument displays the word Overload on front panel and returns "9.9E37" from the
remote interface.