41 SDM3055/SDM3055A
2.6.7 CONFigure:TEMPerature [{RTD|THER|DEFault}
[,{<type>|DEFault}]]
Sets all measurement parameters and trigger parameters to their default values in On-off
mode.
Parameter
Typical Return
<probe_type>
:
{RTD|THER}
,
default THER. The default
command can only choose the built-in sensormanufacturers,
does not support user defined sensor selection.
none
<type>
:
PT100
(for
RTD
only
choice)
or
{BITS90|EITS90|JITS90|KITS90|NITS90|RITS90|SITS90|TITS9
0} (for THER can choose the sensor)
Configuration RTD measurement. Then read measurement results:
CONF:TEMP RTD,PT100
READ?
Typical Response
:
-2.0002
To change temperature units, use UNIT:TEMPerature.
If the input signal is greater than can be measured on the specified manual range, the
instrument displays the word Overload on front panel and returns "9.9E37" from the
remote interface.
Use READ? or INITiate to start the measurement.