![Siemens SIWAREX WP251 Manual Download Page 158](http://html.mh-extra.com/html/siemens/siwarex-wp251/siwarex-wp251_manual_374538158.webp)
Scale parameters and functions
8.36 DR 46 Protocol content
SIWAREX WP251
158
Manual, 12/2015, A5E37203357A
Variable
Note
Type
Length
(bytes)
Rw Default
Min. Max.
Modbus
regis-
ters
String header for
time
String header for time
UBYTE[2] 2
r
8,8
4380
(Page 159)
Time
CHAR[8] 8
r
"23:59:59"
4381
String header for
additional string
String header for addi-
tional string
UBYTE[2] 2
r
2,2
4385
Additional string
Currently not used
CHAR[4] 4
r
" "
4386
Reserve
Reserve
USHORT 2
r
0
4388
Reserve
Reserve
FLOAT
4
r
0
4390
8.36.2
Oldest report ID
The ID of the first saved log is displayed here.
8.36.3
Newest report ID
The ID of the last saved log is displayed here.
8.36.4
Selected protocol-ID
The ID of the log requested in data record 45 and shown in data record 46 is displayed here.
8.36.5
Protocol
The 46 bytes of the log have the following structure:
Designation
Data type
Length
Gross/Net
UBYTE
2
SEMICOLON
UBYTE
1
G/N weight calibratable
UBYTE
8
SEMICOLON
UBYTE
1
Weight unit
UBYTE
4
SEMICOLON
UBYTE
1
Tare sign
UBYTE
2
SEMICOLON
UBYTE
1
Tare
UBYTE
8
SEMICOLON
UBYTE
1
Summary of Contents for SIWAREX WP251
Page 28: ...Mounting 5 3 Mounting on the SIMATIC S7 1200 SIWAREX WP251 28 Manual 12 2015 A5E37203357A ...
Page 212: ...Technical specifications 13 3 Approvals SIWAREX WP251 212 Manual 12 2015 A5E37203357A ...
Page 218: ...List of abbreviations B 1 List of abbreviations SIWAREX WP251 218 Manual 12 2015 A5E37203357A ...