6
OPERATION
54
SITRANS FX330
www.siemens.com/flow
05/2016 - A5E36751269-AA
The table below lists the access levels, their default passwords and associated permissions.
Reset passwords
Reset passwords
Reset passwords
Reset passwords
In case the user has changed particular passwords from their default settings and is not able to
log in anymore, there is a "Reset Passwords" command available in the C6.2.3 submenu.
However to prevent unauthorised use this command itself is protected by a non-changeable
unique password available by contacting the manufacturer.
Access level
Access level
Access level
Access level
Default password
Default password
Default password
Default password
Permissions
Permissions
Permissions
Permissions
User
0000
(any unassigned
password)
•
View device information
•
Configure the display (C5), including changing the display
language and the content of the measurement pages
Operator
0009
•
All rights of "User" access level
•
Configure binary output (C2.2)
•
Configure all HART
®
communication options (C3) with the
exception of "C3.1.1 Current Loop Mode"
•
Change "Operator" password (C6.2.2)
–
note that the new
password must have three leading zeroes ("000")
•
Activate a different meter type
Expert
0058
•
All configuration rights, especially process setup (C1) and
current output (C2.1)
•
Change "Expert" password (C6.2.2)
–
note that the new
password must have two leading zeroes ("00")
INFORMATION!
When a particular access level has been enabled, please pay close attention to a "lock" symbol
which might be present at the right side of a menu entry. This lock symbol indicates that your
current access level allows you to read but not write this entry.
INFORMATION!
When entering the four digit password please be sure to type in all leading zeroes
–
i.e. enter the
default "Operator" password as "0009" and not only "9" by skipping the remaining digits.
INFORMATION!
After a cold start of the device the user role will always be reset to "User".