3
INSTALLATION
32
SITRANS FX330
www.siemens.com/flow
05/2016 - A5E36751269-AA
3.8.4 Mounting the field housing, remote version
Pipe mounting
1
Fix the signal converter to the pipe.
2
Fasten the signal converter using standard U-bolts and washers.
3
Tighten the nuts.
Wall mounting
1
Prepare the holes with the aid of the mounting plate.
2
Use the mounting material and tools in compliance with the applicable occupational health
and safety directives.
3
Fasten the housing securely to the wall.
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
Figure 3-14: Pipe mounting of the field housing
Figure 3-15: Wall mounting of the field housing
INFORMATION!
Signal converters with a wall mounting rack have to be mounted with screws (Ø8 mm / 0.3") or
with U-brackets (Ø8 mm / 0.3") in case of pole installation. In case of mounting directly to the
wall, a mounting system with a minimum load force of 0.1 kN (for example FISCHER type UX10)
suitable for the background has to be applied.