4.6 Ingress protection
The signal converter electronics housing meets the requirements for IP66/67 in accordance with
EN 60529 both for the compact and for the remote version.
Therefore it is essential to observe the following points:
•
Use only original gaskets. They must be clean and free of any damage. Defective gaskets must
be replaced.
•
The electrical cables used must be undamaged and must comply with regulations.
•
The cables must be laid with a loop
1
upstream of the measuring device to prevent water
from getting into the housing.
•
The cable feedthroughs
2
must be tightened. Note that the clamping range of the cable
feedthrough corresponds to the outer diameter of the cable.
•
Align the measuring device so that the cable feedthrough is never facing up
3
.
•
Close any unused cable feedthroughs using blind plugs
4
suitable for the protection category.
•
Do not remove the required cable bushing from the cable feedthrough.
CAUTION!
After all servicing and maintenance work on the measuring device, the specified ingress
protection category must be ensured again.