Patients with an offset heart axis (e.g. dilatative cardiomyopathy)
may require a different orientation than parallel to the spine.
1
Check the expiration date of disposable electrodes and order new
ones if necessary.
2
Pull the protective foil off the electrodes and attach them.
3
Use the application cushion and position it together with the PERU
on the patient. (
Page 89 Preparing the measurement)
4
Connect the electrode clips of PERU to the ECG electrodes.
◆
Perform the examination. See:
Software operator manual
1
Check to see if the leads show a preferably pronounced R-wave on
the Dot display. Information about the ECG signal quality is also
available at the display.
2
In case you are using the Body 18 coil, position it over the heart.
Connect the coil and secure it with the belts. See:
Coil operator
manual
The display shows the following message: Initial Learningphase
active. Don´t move table.
3
Wait at least 10 heart beats (learning phase for triggering) before
you move the patient table into the magnet.
The patient must not move during this learning phase. The
patient table is in the Home position.
The red LED (fault) at the PERU is flashing rapidly?
No analyzable signal is available.
◆
Ensure that the ECG electrodes are attached correctly.
Performing the examination
Evaluating the signal quality
4
Physiological imaging
94
MR System | Operator Manual
Print No. M6-04401.621.02.02.02