
✓
The patient is positioned on the tabletop.
✓
The patient table has been moved to measurement height.
WARNING
Laser beam of the laser light localizer!
Eye injury caused by laser beam
◆
Ensure that the operating and adjustment devices as well as
methods given are used as described.
◆
Inform the patient about the laser and request that he keeps
his eyes closed during positioning.
◆
Ensure that helpless patients keep their eyes closed during
the positioning procedure.
1
Press the Laser Light Localizer button on the control unit.
The laser light localizer is switched on. A crosshair is visible directly
below the area.
2
Move the tabletop so that the crosshairs point precisely to the
region of interest.
The slice for measurement is marked. The display shows the
relative tabletop position of the marked slice.
3
If the laser is correctly positioned, press the Center Position button
for one second to move the table into the magnet isocenter.
The tabletop moves to the selected position and the laser shuts off
automatically.
When the table is not moving, the laser light localizer shuts off
automatically after 60 seconds.
MR system components
3
MAGNETOM Skyra
fit
| syngo MR E11
71
Print No. M6-04401.621.02.02.02