Patients may use the squeeze bulb to alert the operating personnel
(patient alert):
◾
Acoustically:
–
Continuous tone over the intercom
–
Brief feedback signal via the patient’s headphones and
loudspeaker in the examination room
◾
Visually:
–
Blinking button on the intercom
–
Message on the Dot display in the examination room
The alert can be reset by pressing the Reset Patient Alert or the Talk
button on the intercom. Alternatively, the alert reset is possible at the
control unit in the examination room.
Patients, for example, sedated patients, who may not be able to
alert the personnel must be monitored by a person present in the
examination room.
Operating the intercom
The intercom operation is partially software-based. The Patient
Comfort Configuration dialog box allows for several settings, for
example, Active Noise Cancellation to reduce background noise
from the system, can be activated. For detailed information regarding
the operation of the software, see:
Software operator manual
Button
Function
Listen
Enables listening to the patient in the exami-
nation room
Talk
Allows speaking to the patient (as long as the
button is pressed)
Reset Patient
Alert
Resets the patient alert
Physio Signal
Switches the transmission of physiological sig-
nals on/off
Patient alert