
4
Use a dampened cloth for cleaning.
Do not submerge the receptors in cleaning liquid.
5
Disinfect the receptors with a commercially available disinfecting
agent. Follow the manufacturer’s instructions.
Disinfecting system components
Disinfecting sprays damage electronic components. For this
reason, components may only be cleaned with a semi-dry cloth.
WARNING
Flammable cleaning or disinfection agents may cause fire or
explosion!
Injury to patient
◆
Disinfect the system components with commercial
disinfecting materials. However, do not use solutions with
alcohol or acetone. Follow manufacturer’s instructions.
Siemens does not test alcohol-containing, phenol-alkaline and
alkaline-based disinfectants for harmful effects on surfaces.
Avoid disinfecting system components with these materials.
Care and cleaning of floors
Do not use the following cleaning or care products:
◾
Sprays
◾
Silicon-based cleaning or care products
◾
Cleaning or care products with substances that release ammonia
◾
Cleaning or care products that destroy the anti-static properties of
the floor covering
6.1.7
6.1.8
Maintenance
6
MAGNETOM Skyra
fit
| syngo MR E11
119
Print No. M6-04401.621.02.02.02