Application examples
C.3 Example 2: Horizontal vessel with volume measurement
SITRANS Probe LR with mA/HART
Operating Instructions, 05/2019, A5E32337711-AC
97
Parameter
Description
Value
Function
P838
Auto False-Echo Sup-
pression Distance
Distance
1)
minus 0.5 m
Sets length of learned TVT curve
2)
to use
P837
Auto False-Echo Sup-
pression
2, then 1
Enables the use of learned TVT curve
2)
1)
Distance to product from reference point.
2)
For more details on setting P837 and P838, see P837 Auto false echo suppression
(Page 66). For an explanation, see Auto false-echo suppression (Page 88).