Parameters
back
Serial Number
Parameter Device
Software Revision
Configuration
Parameters
back
Configuration
Configuration
0212
Figure 53: Viewing the device configuration
9.4.14
Service
Menu item "Service" serves to retrieve and change the device calibration data set.
Measuring
T=150 °C
Measuring T=150°C
0
1000
CO
564
ppm
Parameters
back
C1:
0.0712
Calibration Values
C2:
0.0712
C3: 500.1234
Parameters
back
Settings
Device
Parameters
Service
Password
0 0 0 0
back
select
Password:
C4: 20.1234
C5:
0.0123
C6:
1.0000
C7:
0.0123
C8:
1.0000
Edit: Enter
par
Figure 54: Calling up the device calibration data
9.4.14.1
Changing the device calibration parameters
WARNING
Endangerment of system safety through calibration parameters with incorrect settings
The calibration values can be changed after the password has been entered. This
is only necessary in special cases, e.g., exchanging the sender lamp. Changing the
calibration values without the necessary technical knowledge can lead to measured
value deviations that can endanger safe operation of the plant.
b
Only change the calibration parameters when you have sufficient technical knowl‐
edge of the device calibration.
b
Only pass the password on to authorized persons.
9
MENUS
68
O P E R A T I N G I N S T R U C T I O N S | GM901-02
8008932/16YT/V4-0/2022-02 | SICK
Subject to change without notice