NOTE
To convert from mg/m
3
to ppm: 1 mg/m
3
N = 0.8 ppm
Example:
Example values:
Calculation:
MB = 1500 ppm
TW [ppm
×
m] = MB [ppm]
×
x
×
S [m]
S = 1 m
TW [ppm
×
m]
×
= 1500 [ppm]
×
0.7
×
1 [m]
Test point at 70% of MBx = 0.7
TW [ppm
×
m]
×
= 1050 [ppm
×
m ]
Result:
Test value of the test cell: 1050
×
m
NOTE
Possible deliverable test cell:
•
1600
×
m
•
4000
×
m
•
10000
×
m
7.11.1.2
Carrying out a SPAN test
Prepare the SPAN test,
.
SPAN test menu navigation,
see "Performing a manual SPAN test", page 71
.
1
Attach the test cell holder on the sender.
NOTICE
Do not insert a cell filled with CO yet.
2
To start the SPAN test, press "cal" on the operator panel of the control unit,
"Performing a manual SPAN test", page 71
.
3
Follow the menu instructions.
1 2
Figure 20: GM901-02 with holder for test cell
1
Input compartment for CO cell
2
Holder for test cell
7.11.2
Online SPAN test according to EPA Guidelines
Criterion for EPA-compliant SPAN test
•
GPP (Gas-Permeation-Probe) with
–
Connection for gas inlet
–
Connection for pressure monitoring in the gas measuring chamber
Filling test or zero gas creates an overpressure in the filter element. As soon as the
overpressure is high enough, it presses the exhaust gas out of the filter. This switches
the gas flow through the filter element to the opposite direction. This method serves to
determine the zero point and various other test points when using suitable test gases.