Measuring
T=150 °C
Measuring
T=150°C
0
1000
CO
564
ppm
Parameters
back
Measuring Range
Limit Value
Parameter Settings
Meas. Distance
Temperature
par
Parameters
back
Settings
Device
Parameter
Service
Parameters
back
Substitute :
150°C
External :
Scale
Low
0 °
C
Scale High : 250 °C
Edit: Enter
Input Low:
4,0
mA
Input High
:
20,0
Temperature Input
Yes
:
:
mA
+
Figure 37: Setting parameters for temperature sensor PT1000
9.4.7.1
Changing the default temperature value
Submenu item "Substitute" serves to enter the default temperature value. The measur‐
ing system uses this value as temperature value when:
•
The external temperature sensor fails.
•
The temperature measurement is outside the configured upper limit value. The
device outputs a warning message in this case.
NOTE
Factory setting: 150 °C
Allowable values: ? ... 430 °C
Contact SICK Service when you are not sure which temperature value should be set.
1
Follow the menu path as described,
see "Parameter settings for the external
temperature sensor PT1000", page 59
.
2
Enter the password,
see "Entering the password", page 52
3
Enter the changes.
4
Confirm value with "Enter".
5
Press the "Left arrow" to return to menu item "Parameter Settings".
back
Substitute :
150°C
External :
Scale Low :
0 °C
Scale High :
250 °C
Edit: Enter
Temperature Input
AnaIn
back
Substitute :
150°C
External :
Scale Low :
0 °C
Scale High :
250 °C
Edit: Enter
Temperature Input
AnaIn
back
Substitute :
150°C
External :
Scale Low :
0 °C
Scale High :
250 °C
Edit: Enter
Temperature Input
AnaIn
back
Substitute :
160°C
External :
Scale Low :
0 °C
Scale High :
250 °C
Edit: Enter
Temperature Input
AnaIn
Figure 38: Setting the default temperature setting (example)
9.4.7.2
Setting the use of an external temperature sensor
Submenu item "External" serves to set a "Yes"/"No" switch as to whether an external
temperature sensor (PT1000) is used. The signal runs via an analog input. "Analn" is
shown when the switch is set to "Yes".
1
Follow the menu path as described,
see "Parameter settings for the external
temperature sensor PT1000", page 59
.
2
Enter the password,
see "Entering the password", page 52
3
Enter the changes.
4
Confirm value with "Enter".
5
Press the "Left arrow" to return to menu item "Parameter Settings".
9
MENUS
60
O P E R A T I N G I N S T R U C T I O N S | GM901-02
8008932/16YT/V4-0/2022-02 | SICK
Subject to change without notice