3
Product description
3.1
Product identification
Product name
GM901-02
Device version
CO measuring device with probe
Manufacturer
SICK AG Erwin-Sick-Str. 1 • D-79183 Waldkirch • Germany
Type plates
•
Sender and receiver: Under the optics tube
•
Measuring probe: On the electronics housing
3.2
Product features
•
The in-situ gas analyzer GM901-02 serves for continuous measurement of gas
concentrations in industrial plants.
•
The GM901-02 is an in-situ measuring system which means measuring is done
directly in the gas carrying duct.
•
Measuring components: CO and reference value temperature.
•
Measuring principle: Infrared spectroscopic gas filter correlation.
3.2.1
Cross-sensitivities
Accurate temperature input is a critical factor that may be required to maintain desired
measurement accuracies. The temperature of an external RTD temperature sensor
should be connected to the analog input of the control unit.
The influence of temperature can be assessed in the following Table.
Process temperature
Absolute temperature
error
Relative temperature
error
Additional relative %
measurement error
100 °C
5 °C
5%
3%
200 °C
10 °C
5%
4.8%
300 °C
15 °C
5%
6%
400 °C
20°C
5%
6.8%
3
PRODUCT DESCRIPTION
12
O P E R A T I N G I N S T R U C T I O N S | GM901-02
8008932/16YT/V4-0/2022-02 | SICK
Subject to change without notice