Chapter
2: Model A Operating
DATOSPIR MICRO User’s Manual
48
511-A00-MU2 • REV. 2.06
FVC X.XX YY%
FEV1: x.xx yy%
X.XX: Observed Value of the parameter, in this case FVC.
YY: % between the Observed value and the value of
Reference.
x.xx: Observed Value of the parameter, in this case FEV1.
yy: % between the Observed value and the value of
Reference.
ATTENTION: Check in the Customisation Programme that
the REFERENCES and the ETHNIC FACTOR are suitably
selected. The ethnic factor modifies the value of the
References according to the percentage selected. 100%
is equivalent to the unmodified standard value of the
References.
Using the selection keys or it is
possible to view the observed value and the percentage in relation
to the value of reference for each customised parameter.
If a parameter does not have a Value of Reference, the
corresponding percentage will not appear.
6
Press the
key again and complete a new
manoeuvre. As many manoeuvres as required can be completed.
The DATOSPIR MICRO will always save the five best for FVC and
VC and the three best for MVV, according to ATS/ERS criteria
(*).
No more than eight manoeuvres should be made so as not to
tire the patient.
Summary of Contents for DATOSPIR MICRO A
Page 1: ...DATOSPIR MICRO 511 A00 MU2 REV 2 06 2019 03 USER MANUAL SPIROMETER...
Page 6: ...Index DATOSPIR MICRO User s Manual 6 511 A00 MU2 REV 2 06...
Page 8: ...DATOSPIR MICRO User s Manual 8 Declaration of Conformity 511 A00 MU2 REV 2 06...
Page 14: ...Safety DATOSPIR MICRO User s Manual 14 511 A00 MU2 REV 2 06...
Page 64: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 64 511 A00 MU2 REV 2 06 U...
Page 67: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 67 511 A00 MU2 REV 2 06 U...
Page 127: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 127 511 A00 MU2 REV 2 06 U...
Page 131: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 131 511 A00 MU2 REV 2 06 U...
Page 133: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 133 511 A00 MU2 REV 2 06 U...
Page 137: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 137 511 A00 MU2 REV 2 06 U...
Page 192: ...Chapter 5 Technical Specifications DATOSPIR MICRO User s Manual 192 511 A00 MU2 REV 2 06...
Page 198: ...Chapter 6 Operating Principles DATOSPIR MICRO User s Manual 198 511 A00 MU2 REV 2 06...
Page 218: ...Appendix 1 Electromagnetic Compatibility DATOSPIR MICRO User s Manual 218 511 A00 MU2 REV 2 06...