Chapter 6: Operating Principles
DATOSPIR MICRO User’s Manual
195
511-A00-MU2 • REV. 2.06
6.3 MICROPROCESSOR
PHYSICAL DESCRIPTION
The microprocessor system is made up of a series of electronic
devices that save, manage, receive and send data. In general
terms, it is divided into:
- Basic hardware control programme (BIOS) residing in the
internal Flash of the Micro controller (128 KBytes).
- Spirometry and equipment management programme and test
database residing in the 2 MByte FLASH memory.
- Non-volatile 512 KByte RAM memory to save the equipment
configuration, status variables and calibration database.
- Central Processing Unit (CPU).
- Communications controller (Series, USB and Bluetooth) with
the outside.
- Clock - Calendar - Alarm.
PROGRAMME
The control programme has been developed in assembler and
in top level C language to ensure very fast time control and a
structured programme. It is divided into two parts: the Bios in
internal Flash and the application in external Flash.
MEMORY
The storage capacity for temporary data, for the customised
equipment configuration and for the calibration database is
512 KB in non-volatile RAM. The test database has a maximum
capacity of 1 MByte.
Summary of Contents for DATOSPIR MICRO A
Page 1: ...DATOSPIR MICRO 511 A00 MU2 REV 2 06 2019 03 USER MANUAL SPIROMETER...
Page 6: ...Index DATOSPIR MICRO User s Manual 6 511 A00 MU2 REV 2 06...
Page 8: ...DATOSPIR MICRO User s Manual 8 Declaration of Conformity 511 A00 MU2 REV 2 06...
Page 14: ...Safety DATOSPIR MICRO User s Manual 14 511 A00 MU2 REV 2 06...
Page 64: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 64 511 A00 MU2 REV 2 06 U...
Page 67: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 67 511 A00 MU2 REV 2 06 U...
Page 127: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 127 511 A00 MU2 REV 2 06 U...
Page 131: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 131 511 A00 MU2 REV 2 06 U...
Page 133: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 133 511 A00 MU2 REV 2 06 U...
Page 137: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 137 511 A00 MU2 REV 2 06 U...
Page 192: ...Chapter 5 Technical Specifications DATOSPIR MICRO User s Manual 192 511 A00 MU2 REV 2 06...
Page 198: ...Chapter 6 Operating Principles DATOSPIR MICRO User s Manual 198 511 A00 MU2 REV 2 06...
Page 218: ...Appendix 1 Electromagnetic Compatibility DATOSPIR MICRO User s Manual 218 511 A00 MU2 REV 2 06...